Self-contained on-chip fluid actuation for flow initiation in liquid cell transmission electron microscopy - GREYC electronique Accéder directement au contenu
Article Dans Une Revue Japanese Journal of Applied Physics Année : 2019

Self-contained on-chip fluid actuation for flow initiation in liquid cell transmission electron microscopy

Résumé

Liquid cells for transmission electron microscopy enable the observation of liquid samples with nanometer resolution, but sample activity must often be triggered via external stimuli including liquid flow and mixing. Existing devices enable flow using specialized TEM holders and external pumping setups which limits technique accessibility. Here, a liquid cell with integrated MEMS valves and pumps is demonstrated with flow capabilities using an electrical-contact holder. On-chip electrochemical pumps and capillary burst valves allow for the injection of fluid from an integrated reservoir into an observation chamber using only electrical signals from a general microchip-sized TEM holder.
Fichier principal
Vignette du fichier
Menon_2019_Jpn._J._Appl._Phys._58_090909.pdf (813.69 Ko) Télécharger le fichier
Origine Fichiers éditeurs autorisés sur une archive ouverte

Dates et versions

hal-02335385 , version 1 (28-05-2024)

Licence

Identifiants

Citer

Vivek Menon, Matthieu Denoual, Hiroshi Toshiyoshi, Hiroyuki Fujita. Self-contained on-chip fluid actuation for flow initiation in liquid cell transmission electron microscopy. Japanese Journal of Applied Physics, 2019, 58 (9), pp.090909. ⟨10.7567/1347-4065/ab386a⟩. ⟨hal-02335385⟩